The comparison of capnography and epigastric auscultation to assess the accuracy of nasogastric tube placement in intensive care unit patients

Abstract

Background

Placement of nasogastric (NG) tubes is a common procedure for patients especially in intensive care units (ICUs). Thus, it is important to determine the correct placement of the tube to prevent misplacement in the airways. Accordingly, the aim of this study was to compare the epigastric auscultation and capnography in assessing the accuracy of NG tube insertion in ICU patients.

Methods

In this descriptive comparative study, 60 patients were selected trough convenience sampling. After insertion of the NG tube in a standard method, the accuracy of placement of the tube with both epigastric auscultation and capnography was investigated. The NG tube insertion accuracy was then confirmed via radiography. Data analysis was performed using statistical software SPSS version 23.

Results

The result showed that capnography had a sensitivity, specificity, and accuracy of 100, 92.5, and 95% respectively, but epigastric auscultation had 90, 80, and 83.4% respectively. The Kappa agreement coefficient between two methods was − 0.759.

Conclusion

The results revealed that the use of the capnography is preferable over the epigastric auscultation to confirm the correct insertion of the NG tube. It is recommended that more than one method be applied to detect and confirm the correct insertion of the NG tube.

Peer Review reports

Background

Patients admitted to the emergency department and the intensive care units (ICUs) with a reduced level of consciousness in general often require nasogastric (NG) tube insertion [1]. The NG tubes are widely used in special sectors, with the annual application of this equipment being 1.2 million in the United States [2] and more than 1 million in the UK [3]. In general, NG tube insertion is carried out with various diagnostic and treatment purposes [4]. Typically, one of the most important goals of using these tubes is to supply nutritional needs [5]. Other applications of these tubes include drug therapy and reducing stomach pressure, especially in abdominal operations [6].

Despite all the benefits of NG tubes, their inappropriate placement could endanger the safety of patients [7]. Most insertions are unsafe and lead to the accidental entry of NG tubes into airways and routes other than the gastrointestinal tract [3, 8]. Airway misplacement of NG tubes is the most common issue in this regard [9]. With the incorrect placement of NGs, about 3600–8400 cases of lung injuries and 1200–3600 deaths occur in the United States on an annual basis [2]. Misplacement of NG tubes causes serious problems, including aspiration, pneumothorax, pneumonia, bronchopleural fistula, emphysema, pulmonary hemorrhage, mediastinal inflammation, esophageal perforation, atelectasis, and mortality [10].

Today, various clinical methods are employed to determine the location of the NG tube, some of which are not reliable [11] or not available at the patient’s bedside. These methods include measuring the bilirubin concentration, pepsin, and trypsin of aspirated materials [12]. A new method for locating NG tubes in patients is measuring the carbon dioxide output from the expiratory flow of the lungs, which is carried out by a device called capnograph [13]. Since carbon dioxide exists only in the expiratory air, capnography can be applied to distinguish the airway from the gastrointestinal tract in the tube placement= [14].

In a study, this method was introduced as a safe and cost-effective approach in evaluating the NG tube, which promotes patient care [15]. Meanwhile, another research considered this approach ineffective in this regard [16]. Some sources have described the use of a capnograph along with another method to ensure the correct insertion of the NG tube [15]. Another method for determining and confirming the location of the tube in patients is the epigastric auscultation, which involves instilling air into the feeding tube while using a stethoscope placed over the stomach. Some of the benefits of this technique include simplicity and cost-effectiveness [17, 18]. Nevertheless, different sources consider this traditional technique as an unreliable test [5]. While epigastric auscultation is most commonly applied by nurses [19] and is introduced as an unreliable traditional technique in some sources [20, 21], its superiority over capnograph has been shown in some studies [22, 23]. Thus, considering the failure to identify the desirable method in previous studies, this study aimed to compare capnography and epigastric auscultation in confirming the correct placement of the NG tube.

Methods

This was a descriptive and comparative study performed in two intensive care units (ICUs) of hospitals of Kermanshah, Iran in June 2018 to June 2019 (The ICUs in the hospitals affiliated to Kermanshah University of Medical Sciences were evaluated). A total of 60 patients were selected through convenience sampling, and the sample size was estimated using the sample size formula along with the formula for estimating the ratio of a qualitative attribute with a 95% confidence level for all patients intubated.

To determine the sample size using the estimation formula of the ratio of a qualitative trait in the study population and with 95% confidence (α-1) and based on the study by Alpieren et al. [16]. Diagnosis of gastric tube insertion as well as the ratio of gastric nasal tube placement correctly diagnosed as gastric epigastric insertion were calculated as 0.95 below.

Sample Size Determination Based on Ayten Zaybak:

Gastric nasal tube placement ratio diagnosed correctly by gastric capnometry (sidestream) 0.95 and gastric nasal tube placement ratio diagnosed correctly by gastric epigastric method 0.90 is calculated as follows.

Research tools included demographic characteristics form (age, gender, disease diagnosis, and admitted hospital), physiological information checklist (blood pressure, pulse, and arterial oxygen saturation), and Glasgow Coma Scale, which is designed to evaluate the level of consciousness and responses of patients to stimuli which is carried out based on three behaviors of patients. The participants were enrolled in the study after selecting the target group and obtaining written consent from their legal guardians. The questionnaires were completed by the first author. Before the start of the experiment, the radiologist was informed to attend the patient’s bedside. The NG tube was inserted by the relevant nurse in a standard manner. The location of the tube was evaluated using the epigastric auscultation. In this regard, after the completion of the intubation, 10 ml of air was blown into the tube through a syringe attached to the end of the NG tube. Simultaneously, the air flow was listened to by the Littmann classic II S.E. Once the NG tube is impregnated with lidocaine gel, it is directed from one of the nasal passages to the esophagus, then to the stomach, and then to the NG fixes with the CM scalp.

Interventions for reducing the leak from the connection between NG tube and calorimetric capnography (sidestream) probe were done with connection. Standard CO2 levels were used for gastric ultrasound [24]. Stethoscope (made in the USA), and the presence or absence of the sound of the air motion was recorded.

$$ {\displaystyle \begin{array}{l}\alpha =0.05\to {Z}_{1-\raisebox{1ex}{$\alpha $}\!\left/ \!\raisebox{-1ex}{$2$}\right.}=1.96\\ {}P=0.95\\ {}d=0.1\\ {}n=\frac{{\left({z}_{1-\raisebox{1ex}{$\alpha $}\!\left/ \!\raisebox{-1ex}{$2$}\right.}\right)}^2P\left(1-P\right)}{(d)^2}=\frac{(1.96)^20.95\left(1-0.95\right)}{(0.1)^2}=\frac{(3.84)(0.0475)}{(0.1)^2}=\frac{0.1824}{0.01}=18.24\\ {}n\ge 18\\ {}\\ {}\alpha =0.05\to {Z}_{1-\raisebox{1ex}{$\alpha $}\!\left/ \!\raisebox{-1ex}{$2$}\right.}=1.96\\ {}P=0.90\\ {}d=0.1\\ {}n=\frac{{\left({z}_{1-\raisebox{1ex}{$\alpha $}\!\left/ \!\raisebox{-1ex}{$2$}\right.}\right)}^2P\left(1-P\right)}{(d)^2}=\frac{(1.96)^20.90\left(1-0.90\right)}{(0.1)^2}=\frac{(3.84)(0.09)}{(0.1)^2}=\frac{0.3456}{0.01}=34.56\\ {}n\ge 35\\ {}\\ {}\\ {}\end{array}} $$

Next, a capnograph adapter (Alborz B9 vital signs monitor, made in Iran) was connected to the end of the NG tube and the result was recorded after 1 min. In the final stage, chest radiography was performed to determine the exact location of the tube. The results of the other two techniques were confirmed or rejected according to the radiographic diagnosis. Data analysis was performed in SPSS version 23 using descriptive statistics (frequency distribution, mean and standard deviation of the variables) to estimate the sensitivity, specificity, and accuracy of capnography and epigastric auscultation techniques, and to compare the accuracy of the two methods.

Results

The mean age of patients was 47.06 ± 19.11. Also, 83.3% of the patients were male and 95% of patients were admitted to the ICU; 35% had systolic blood pressure less than 120 and 48.3% had systolic blood pressure above 120 mmHg. Further, 45% had diastolic blood pressure less than 80 and 43.3% had diastolic blood pressure above 80 mmHg. Finally, 13.3% had a pulse rate below 60 and 63.3% had a pulse rate above 80 beats / min (Table 1).

Table 1 Demographic characteristics of participants

The results revealed that the sensitivity, specificity and accuracy of the capnography were higher than those of the epigastric auscultation. Capnographic sensitivity of 100%, specificity of 92.5%, and positive and negative predictive values of 86.9 and 100% were reported at 95% accuracy (Table 2).

Table 2 sensitivity, specificity, PPV, NPV and accuracy of epigastric auscultation and capnography in assessing the nasogastric tube placement in the ICU patients

The Kappa agreement coefficient was used to determine the agreement between two methods in the correct placement of the gastric tube. In terms of agreement between variables of capnography)respiratory placement of nasogastric tube(and epigastric auscultation)respiratory placement of nasogastric tube(, the percentage of agreement was 80.8%. Also, for variables of capnography) gastric placement of nasogastric tube(and epigastric auscultation) gastric placement of nasogastric tube (, the agreement was 94.1%; according to Kappa agreement, significant results were obtained (p < 0.05) (Table 3).

Table 3 Kappa agreement coefficient of epigastric auscultation and capnography in assessing the nasogastric tube placement in the ICU patients

The area under ROC curve of capnography and epigastric auscultation showed a better performance of capnography (Fig. 1).

Fig. 1
figure 1

Area under ROC curve of capnography and epigastric auscultation

Discussion

According to the results of the current study, capnography had a higher sensitivity, specificity, and accuracy, compared to the epigastric auscultation technique. In a research, Kindopp et al. evaluated capnography in determining feeding tube insertion in ICU patients. According to the results, the use of capnography for determining the location of the NG tube inserted in the airway had 100% sensitivity and specificity. Thus, capnography can be employed in placing the NG tube to increase patient safety [25]. In another study, Haghighi Moghadam et al. [26]. noted that capnography had 100% accuracy in distinguishing the airway from the gastrointestinal tract, where capnography results had 100% correlation with radiography results, the golden standard for determining the location of the NG tube. As such, capnography can be used as a substitute for radiography to distinguish the gastrointestinal tract from the airway during NG tube insertion [1]. In another study by Bruns et al., capnography was applied to determine the location of the NG tube as a simple, low-cost, and reliable method [5].

While capnography results are appropriate in studies, they are presented as false positive when the NG tube is left in the mouth or the throat. Some studies also confirm this claim [12, 25], and if the tube does not enter the stomach sufficiently and remains in the esophagus, the capnography will show the results as false positive, in which case the risk of aspiration will increase [5]. Our findings are somehow different from similar studies. In a study by Rahimi et al., the diagnostic percentage of epigastric auscultation was 33.3% in the detection of tube insertion in the lung and 29.2% in gastric tube insertion. These scholars expressed that despite the widespread use of epigastric auscultation, it is not a reliable technique [6]. In spite of a low detection of feeding tube insertion with epigastric auscultation, this method is used relatively frequently. In a research by Roynette et al., which was performed in 383 ICUs of 20 European countries, the results demonstrated that 84.7% of nurses applied the epigastric auscultation method [27]. In another study, the results indicated that 86% of nurses employed the epigastric auscultation technique to confirm the feeding tube insertion [28].

Some studies have evaluated both epigastric auscultation and capnography in confirmation of the feeding tube insertion. In a research by Meyer et al., the accuracy of a combined method of epigastric auscultation and colorimetric capnography was evaluated in confirming the correct location of the NG tube in ICU patients connected to a mechanical ventilator. Carbon dioxide was observed in 9 out of 69 cases of NG tube insertion, two and seven of which were placed in the airway and the stomach, respectively. Nevertheless, the findings indicated a complete correlation between colorimetric capnography and epigastric auscultation for confirming the location of the NG tube. Further, the present research revealed that colorimetric capnography along with epigastric auscultation is a safe and low-cost method that promotes health care for patients [29]. In a study by Galbois et al., the results indicated that the use of capnography and epigastric auscultation reduced the costs and time required for determining the location of the NG tube, and that these are safe and cost-effective techniques for improving nursing care [15].

In a previous research, Erzincanli et al. evaluated colorimetric capnometry in confirming the NG tube insertion, reporting a 95% correlation between colorimetric capnography and radiography and an 82.5% correlation between epigastric auscultation and radiography. Furthermore, the results of the mentioned study revealed 100% sensitivity and 66.7% specificity of colorimetric capnography in confirming the insertion of the NG tube. While colorimetric capnography has been successful in the distinguishing of the airway from the respiratory tract, it fails to separate the tube inserted in the stomach and/or duodenum [22]. Alpiren et al. conducted a research to compare the accuracy of capnometry and epigastric auscultation in the placement of NG tube to determine the correct place of NG tube in ICU patients. These scholars reported that capnometry detected wrong location in 15 (16%) out of 91 cases of tube insertion (lung instead of stomach), while epigastric auscultation detected the wrong location in five (5%) out of 91 cases of tube insertion (lung instead of stomach), which is inconsistent with our findings. The results obtained demonstrated that neither epigastric auscultation nor capnometry were reliable methods for confirming the NG tube insertion, and chest radiography was the only reliable technique for determining the accurate location of the tube [1]. This lack of consistency between the results might be due to neglecting the time for removal or absorption of carbon dioxide present in the stomach. In the current research, 1 min was considered for absorption or removal of carbon dioxide in the stomach through the NG tube.

Limitations

The limitations of this study were the low sample size and low level of consciousness of patients, which can reduce the generalizability of findings.

Conclusion

According to the results, both techniques are simple and inexpensive, but the use of capnography is more preferred over epigastric auscultation. However, none of these methods can absolutely predict the correct insertion of the NG tube. Thus, it is recommended that more than one method be applied to detect and confirm the insertion of the NG tube and a combination of both capnography and epigastric auscultation.

Availability of data and materials

Datasets are available through the corresponding author upon reasonable request.

References

  1. Rivera R, Campana J, Hamilton C, Lopez R, Seidner D. Small bowel feeding tube placement using an electromagnetic tube placement device: accuracy of tip location. J Parenter Enter Nutr. 2011;35(5):636–42.

    Article  Google Scholar 

  2. Krenitsky J. Blind bedside placement of feeding tubes: treatment or threat? Pract Gastroenterol. 2011;35(3):32–42.

  3. Cole E. Improving the documentation of nasogastric tube insertion and adherence to local enteral nutrition guidelines. BMJ Qual Improv Rep. 2015;4(1):u203207–w1513.

  4. Mandal M, et al. Nasogastric tube placement--a simple yet difficult procedure--a review. J Evol Med Dent Sci. 2017;6(31):2572–7.

    Article  Google Scholar 

  5. Brun PM, Bessereau J, Cazes N, Querellou E, Chenaitia H. Lung ultrasound associated to capnography to verify correct endotracheal tube positioning in prehospital. Am J Emerg Med. 2012;30(9):2080.e5–6.

    Article  Google Scholar 

  6. Rahimi M, Farhadi K, Ashtarian H, Changaei F. Confirming nasogastric tube position: methods & restrictions: a narrative review. JNMS. 2015;2(1):55–62.

    Article  Google Scholar 

  7. Araujo-Preza CE, Melhado ME, Gutierrez FJ, Maniatis T, Castellano MA. Use of capnometry to verify feeding tube placement. Crit Care Med. 2002;30(10):2255–9.

    Article  Google Scholar 

  8. Giantsou E, Gunning KJ. Blindly inserted nasogastric feeding tubes and thoracic complications in intensive care. Health. 2010;2(10):1135.

    Article  Google Scholar 

  9. Anziliero F, Corrêa AP, Silva BA, Soler BE, Batassini É, Beghetto MG. Nasoenteral tube: factors associated with delay between indication and use in emergency services. Rev Brasileira de Enfermagem. 2017;70(2):326–34.

    Article  Google Scholar 

  10. Bennetzen LVHS, Larsen P. Diagnostic accuracy of the methods carried out to verify nasogastric tube position in mechanically ventilated adult patients: a systematic review protocol. JBI Database System Rev Implement Rep. 2013;11(12):102–9.

    Article  Google Scholar 

  11. Ellett MLC. What is known about methods of correctly placing gastric tubes in adults and children. Gastroenterol Nurs. 2004;27(6):253–9.

    Article  Google Scholar 

  12. Metheny NA, Meert KL, Clouse RE. Complications related to feeding tube placement. Curr Opin Gastroenterol. 2007;23(2):178–82.

    Article  Google Scholar 

  13. Butler BD, Little T, Drtil S. Combined use of the esophageal-tracheal Combitube with a colorimetric carbon dioxide detector for emergency intubation/ventilation. J Clin Monit. 2005;11(5):311.

    Article  Google Scholar 

  14. Hanrahan K, Wagner M, Matthews G, Stewart S, Dawson C, Greiner J, et al. Sacred cow gone to pasture: a systematic evaluation and integration of evidence-based practice. Worldviews Evid-Based Nurs. 2015;12(1):3–11.

    Article  Google Scholar 

  15. Galbois A, Vitry P, Ait-Oufella H, Baudel J-L, Guidet B, Maury E, et al. Colorimetric capnography, a new procedure to ensure correct feeding tube placement in the intensive care unit: an evaluation of a local protocol. J Crit Care. 2011;26(4):411–4.

    Article  Google Scholar 

  16. Elpern EH, Killeen K, Talla E, Perez G, Gurka D. Capnometry and air insufflation for assessing initial placement of gastric tubes. Am J Crit Care. 2007;16(6):544–9.

    Article  Google Scholar 

  17. Farrington M, Lang S, Cullen L, Stewart S. Nasogastric tube placement verification in pediatric and neonatal patients. Pediatr Nurs. 2009;35(1):17.

    PubMed  Google Scholar 

  18. Chan E-Y, Ng IH-L, Tan SL-H, Jabin K, Lee L-N, Ang C-C. Nasogastric feeding practices: a survey using clinical scenarios. Int J Nurs Stud. 2012;49(3):310–9.

    Article  Google Scholar 

  19. Ryu J-A, Choi K, Yang JH, Lee D-S, Suh GY, Jeon K, et al. Clinical usefulness of capnographic monitoring when inserting a feeding tube in critically ill patients: retrospective cohort study. BMC Anesthesiol. 2016;16(1):122.

    Article  Google Scholar 

  20. Simons SR, Abdallah LM. Bedside assessment of enteral tube placement: aligning practice with evidence. AJN the American Journal of Nursing. 2012;112(2):40–6.

    Article  Google Scholar 

  21. Proehl JA, Heaton K, Naccarato MK, Crowley MA, Storer A, Moretz JD, et al. Emergency nursing resource: gastric tube placement verification. J Emerg Nurs. 2011;37(4):357–62.

    Article  Google Scholar 

  22. Erzincanli S, Zaybak A, Güler A. Investigation of the efficacy of colorimetric capnometry method used to verify the correct placement of the nasogastric tube. Intensive Crit Care Nurs. 2017;38:46–52.

    Article  Google Scholar 

  23. Baratloo A, Hosseini M, Negida A, El Ashal G. Part 1: simple definition and calculation of accuracy, sensitivity and specificity. Emerg (Tehran). 2015;3(2):48–9.

    Google Scholar 

  24. Sum Ping ST, Mehta MP, Symreng T. Reliability of capnography in identifying esophageal intubation with carbonated beverage or antacid in the stomach. Anesth Analg. 1991;73(3):333–7.

    Article  CAS  Google Scholar 

  25. Kindopp AS, Drover JW, Heyland DK. Capnography confirms correct feeding tube placement in intensive care unit patients. Can J Anesth. 2001;48(7):705–10.

    Article  CAS  Google Scholar 

  26. Haghighi Moghadam Y, Mohammad Nahal L, Feizi A, Valizadeh HM. A survey on the accuracy of Capnography comparing to chest X-ray in nasogastric tube placement. J Urmia Nurs Midwifery Faculty. 2014;12(3):199–202.

    Google Scholar 

  27. Roynette C, Bongers A, Fulbrook P, Albarran J, Hofman Z. Enteral feeding practices in European ICUs: a survey from the European federation of critical care nursing associations (EfCCNa). E Spen Eur E J Clin Nutr Metab. 2008;3(2):e33–e9.

    Article  Google Scholar 

  28. Persenius MW, Larsson BW, Hall-Lord M-L. Enteral nutrition in intensive care: nurses’ perceptions and bedside observations. Intensive Crit Care Nurs. 2006;22(2):82–94.

    Article  Google Scholar 

  29. Meyer P, Henry M, Maury E, Baudel J-L, Guidet B, Offenstadt G. Colorimetric capnography to ensure correct nasogastric tube position. J Crit Care. 2009;24(2):231–5.

    Article  Google Scholar 

Download references

Acknowledgements

This work was performed in partial fulfillment of the requirements for MS in Critical Care Nursing of Elahe Heidarzadi in Faculty of Nursing and Midwifery, Kermanshah University of Medical Sciences, Kermanshah, Iran and was supported by funding from the Kermanshah University of Medical Sciences (grant no 97109).

Funding

The funding for this research was provided by the deputy of research and technology –Kermanshah University of Medical Sciences, grant No. 97109, the deputy of research and technology –Kermanshah University of Medical Sciences had no role in the design of the study and collection, analysis, and interpretation of data and in writing of the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

R.J. and E.A. equally contributed to the conception and design of the research; R.J. contributed to the design of the research; N.S. contributed to the acquisition and analysis of the data; R. J and B.H. contributed to the interpretation of the data; and E.A. and R. J. drafted the manuscript. All authors critically revised the manuscript, agree to be fully accountable for ensuring the integrity and accuracy of the work, and read and approved the final manuscript.

Corresponding author

Correspondence to Rostam Jalali.

Ethics declarations

Ethics approval and consent to participate

Written informed consent was obtained from all participants in the study and this study approved by Kermanshah University of Medical Sciences ethics committee IR.KUMS.REC.1397.133. https://ethics.research.ac.ir.

Consent for publication

All authors read and confirmed the manuscript. Consent for publication ‘Not Applicable’ for participants.

Competing interests

The authors declare that they have no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Heidarzadi, E., Jalali, R., Hemmatpoor, B. et al. The comparison of capnography and epigastric auscultation to assess the accuracy of nasogastric tube placement in intensive care unit patients. BMC Gastroenterol 20, 196 (2020). https://doi.org/10.1186/s12876-020-01353-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12876-020-01353-5

Keywords