In this study we evaluated the feasibility of using a new device, the LP, and confirmed its safe and effective use in clinical practice when performing third-space procedures (POEM and G-POEM).
Third-space endoscopy was first described by Sumiyama et al. [25]; it was first used in animals in 2007, and then used in humans. It is based on the creation of a submucosal tunnel to perform surgical procedures with confirmed safety and efficacy. It transforms the concept of endoluminal endoscopy to intramural, making many diseases that previously would have been treated by open or laparoscopic surgery endoscopically curable [1, 4].
The lumen is considered as the first space, while the peritoneum is considered the second space and the intentionally-created tunnel is the third space (space between the mucosa and muscularis propria) [2,3,4]. Different disorders have been addressed by this technique, including Zenker’s diverticulum, myotomy for achalasia, gastroparesis, Hirschsprung’s disease, removal of tumors arising from the muscularis propria and beyond, and stricture treatment [1, 2, 8]. Among them, achalasia and gastroparesis are the most prevalent diseases treated by this technique. However, “difficult”, “incomplete”, or even “not possible” tunnel creation or myotomy have been described and are associated with certain complexity factors such as end-stage disease (POEM) [5], previously-treated cases (POEM and G-POEM) [4, 8, 24], inability to identify EGJ (POEM) [6,7,8], PMR (G-POEM) [9], high submucosal fibrosis (POEM and G-POEM) [2,3,4, 6,7,8], lack of experience (POEM and G-POEM) [2, 3, 6],or anatomical factors (POEM) [1, 4, 9, 10]. These factors result in a failed procedure, even if the endoscopist believes that the procedure was successful [1, 3]. If the tunnel is too short, the procedure is ineffective; if the length of the myotomy is too long, there is a higher risk of adverse events, including perforation or bleeding [3, 12]. Currently, there are several endoscopic landmarks, such as palisading vessels at the EGJ and the circular bundle of LES fibers in POEM (difficult use and inaccurate), or the continuous insertion and extraction of the endoscope from the tunnel, that are used to identify the PMR, which is technically challenging. A second endoscope (POEM and G-POEM) [12, 13] and fluoroscopy (G-POEM) [9, 11] are used in an attempt to overcome these problems; however, these methods are costly or are unavailable. Therefore, we decided to explore a new alternative to overcome these problems when performing third-space endoscopy.
The use of lighting devices has been explored in medicine for years. In colorectal and gynecological surgeries, for example, iatrogenic ureteric injury is a serious complication with a variable incidence between 0. 7-10% [14]. The identification of ureters is challenging and the optional double J stent placement is invasive and associated with serious adverse events [15];however, the use of fluorescence and lighted ureteral stents has overcome these problems. LED technology was invented in 1907 by H. J. Round but was commercially available in 1962 in electrical components. Modern LED technology with more practicality was used after 2010 [17]; uses in medicine are confirmed in ophthalmologic procedures where improved illumination for vitrectomy has been observed [16].LED has advantages over incandescent light sources. It provides lower energy consumption, a longer lifetime, smaller size, faster switching, a better spectrum of light and intensity (emitting more lumens per watt compared with light bulbs), and cool light that radiates minimal heat. It is safe because mercury or other hazardous metals are not contained within it [16, 17].
Because of safety and efficiency, we decided to use a white LED-probe and orally insert it into a conventional 127 cm × 20 Fr nasogastric feeding tube. We spent between 10 to 15 min for LP building and disinfection process. Excellent visualization was obtained with the 112 lm/m of the probe; this was enough to be visualized under the submucosal tunnel or over the intraluminal space when inserted into the tunnel. We didn’t have technical problems during assembly or during procedures, neither when insertion into the patient was performed or after LP was used and procedures were finished. Therefore, based on these results, we confirmed the safety and efficacy of this device, that had a median insertion time of 5 min (4-6) for POEM and 6 (5-7) for G-POEM, without compromising total procedural times, and being similar to those observed in previous studies [5,6,7, 10, 11, 21, 24].
POEM was performed as described by other groups [5, 6, 21], with similar demographic characteristics. However, in our cohort 23.8% were previously treated and 38% grade III and IV. These are the subgroups theoretically more difficult to treat; therefore, with the greatest benefit if LP is used. Nonetheless, POEM was completed in 100% of cases and the mean myotomy length was 10.5 ± 3.1 cm, which is similar to the length of 9.4 ± 3.1 cm obtained in other studies [6]. Grimes et al. [13] compared double-scope vs conventional POEM in a clinical trial that included 50 patients per group. No differences in technical (98% vs 100%) or clinical success (93% vs 97%) was found, but with a 34% longer myotomy and 17 min increase in procedural times for double-scope group. In our study, the LP was 6 mm in diameter, which is similar to the length of the neonatal endoscope used in Grimes’ study. However, our LP system was advantageous in terms of cost (10 dls per LP), and placement time (5 min vs 17 min). We confirmed an adequate myotomy in all cases including 6 patients (14.2%) who were considered as difficult (4 grade IV, 1 grade III and 1 pneumatic dilation), in whom classic POEM didn’t complete myotomy and who benefited from the LP use, avoiding potential adverse events or risk of incomplete procedure. Additionally, no other endoscopy tower was needed (saving costs and space in the endoscopy room).. In 2019, Grimes et al. published the follow-up of the cohort of double-scope vs conventional POEM, with a median of 3 years. They found no differences in clinical outcomes between groups (83% vs 80%; P = 1.0), without differences in reflux disease incidence, but more cases with grade B esophagitis were presented in treatment group (25% vs 4%; P = 0.049); they hypothesize that this is because a longer myotomy is performed in them (1.6 ± 1.2cms) [26]. In our cases, the LP allowed performing an adequate EGJ myotomy and, at 6-month evaluation, clinical outcomes and the occurrence of reflux disease was similar to those of other studies [5,6,7,8, 12, 13, 21]. This suggests that the adequate confirmation of EGJ myotomy is the most important step in POEM procedures, regardless of whether an external device is used or not. Confirmation, which should be performed in all cases, mostly in early-experienced endoscopists in POEM procedure, represents LP as an excellent alternative for this purpose.
We performed the G-POEM and LP placement in all cases. Demographic characteristics were similar to other groups [8,9,10,11, 24]. The pylorus was previously manipulated in 21.4% of cases (botulinum toxin injection and transpyloric stent), potentially difficult cases. However, median G-POEM time (60 min) was similar to other groups [1, 9,10,11,12, 24]. Xue H et al. [9] compared the use of fluoroscopy-guided G-POEM vs conventional G-POEM procedure in 14 patients; all procedures achieved technical success, the PMR was identified in all 7 patients of the fluoroscopy-guided group, and only in 4(57.1%) from the control (P < 0.03). However, this was not clinically expressed, with a non-statistically significant difference between GCSI and GES. In our group, LP provided a better orientation towards PMR identification and myotomy confirmation in all cases, including 5/28 patients where the endoscopist was “lost”, during tunnel creation and PMR was not identified, where the LP allowed the completion of G-POEM procedures, representing a 17.8% benefit in them. However, besides the fact that the G-POEM outcomes were slightly better in our study, when compared with other centers, with a general clinical success (85.7% vs 69–81%) in GCSI and GES (89.2% vs 69–84.2%) at 6-month evaluations, we can’t assume that this could be explained because of the 100% PMR identification (similar to the fluoroscopy-guided group from Xue’s study). However, as stated by other authors, different gastroparesis subtypes with their corresponding physiopathology could explain the real heterogeneous mid-term results more than the simple direct effect of the PMR cutting, inclusive with LP guidance, as in our patients [1,2,3,4, 9,10,11, 24].
The strengths of our study include the use of LP in the two most common and important third-space procedures, the sample size that was reached in both and calculated for statistical significance, adequate and strict procedural and follow-up protocols, technical confirmation of all steps in all cases, and excellent safety without adverse events associated with LP use. Our study also has limitations that should be addressed. First, LP is not yet commercially available. Second, LP has to be made before each procedure by the medical doctor, which, in spite of the fact that it takes only between 10 and 15 min, could be time-consuming. Third, different LED and nasogastric feeding tube brands exist around the world, which limits the availability of the system we used. Fourth, only POEM and G-POEM cases were included; pediatric and other third-space procedures were not included, and fifth, LP was used in 76.2% of naïve POEM cases, which represent a subgroup of non-difficult cases in which LP could have been useless, especially when performed by highly-experienced endoscopists in third-space procedures; therefore, we think that the best advantage of LP use could be found in early-experience endoscopists in third-space procedures. However, we think that LP is a useful device for POEM and G-POEM procedures because of its simplicity, innovation, low costs, safety and the ability to make difficult procedures potentially easier.