Effectiveness of introducing a 20-gauge core biopsy needle with a core trap in EUS-FNA/B for diagnosing pancreatic cancer

Abstract

Background

Endoscopic ultrasound-guided fine-needle aspiration/biopsy (EUS-FNA/B) is a standard method for pathological diagnosis of pancreatic solid lesions. The EchoTip ProCore 20G® (PC20), a 20-gauge biopsy needle with a forward-bevel core trap, has been available in Japan since 2015.

Methods

We compared the efficacy of the PC20 with that of the EchoTip ProCore 22G® (PC22) and Acquire 22G® (AC22) in EUS-FNA/B for diagnosing pancreatic cancer. This retrospective study included 191 patients with pancreatic cancer who underwent EUS-FNA/B using the PC20, PC22, or AC22 at our facility from April 2013 to October 2019. We investigated the patients’ clinical characteristics and the diagnostic accuracy and safety of each needle.

Results

A sufficient stroke length of puncture was secured in all patients. The maximum length under EUS was shorter with the AC22 (22.1 ± 2.2 mm) than PC20 (30.6 ± 0.7 mm, p < 0.01) and PC22 (30.3 ± 0.8 mm, p < 0.01). The histological accuracy was 96.4% with the PC20 but only 58.8% with the PC22 (adjusted p (p-adj) < 0.0001) and 75.0% with the AC22 (p-adj = 0.06). The diagnostic accuracy of the combination of histology and cytology was 96.4% with the PC20, while it was 72.1% with the PC22 (p-adj < 0.0001) and 91.7% with the AC22 (p-adj > 0.99). One patient (0.9%) in the PC20 group developed mild pancreatitis, but no adverse events occurred with the other needles.

Conclusions

The PC20 showed better diagnostic capability than the PC22. The diagnostic efficacy was similar between the PC20 and AC22. The high histological accuracy of the PC20 could be advantageous for lesions in which histological assessment is critical.

Peer Review reports

Background

Pancreatic cancer is a highly aggressive tumor with a poor prognosis [1,2,3,4,5]. The age-standardized incidence rate of pancreatic cancer increased by 1.03% per year from 1973 to 2014 [6]. Changes in environmental and behavioral factors rather than genetic factors are thought to be related to the worldwide increase in pancreatic cancer [7, 8]. Differences in lifestyle (e.g., high-calorie diet and high body mass index), exposure to environmental risk factors (e.g., smoking), and accessibility to medical services including more efficient diagnostic tools might be associated with the variability in the mortality of pancreatic cancer among countries [7, 9, 10].

Pancreatic cancer is typically found as a pancreatic solid lesion. Pancreatic solid lesions include pancreatic malignant lesions such as pancreatic cancer, pancreatic neuroendocrine tumors, and metastatic pancreatic cancer as well as benign lesions such as autoimmune pancreatitis and mass-forming pancreatitis. Therefore, prompt and accurate diagnosis of pancreatic cancer is critical to determine the most appropriate therapeutic strategies. For this purpose, pathological diagnosis with minimal invasion is crucial. Endoscopic ultrasound-guided fine-needle aspiration/biopsy (EUS-FNA/B) is the standard method for sampling pathological specimens from a pancreatic solid lesion. The reported diagnostic sensitivity, specificity, and accuracy are 91.7–93.4%, 100%, and 92.3% to 94.4%, respectively [11, 12]. Various factors, including the type of puncture needle, puncture route, number of passes, tumor size, and experience of the endoscopist and cytopathologist, can impact the accuracy of diagnosis [13,14,15]. Several types of puncture needles with various characteristics (e.g., thickness and needle-tip shape) are commercially available [16,17,18]. Also, the needles are categorized into FNA needles and FNB needles and the systematic review by Li et al. [19] presented FNB are superior. The volume of a specimen obtained by core needle biopsy can depend on the thickness of the puncture needle; i.e., a needle with a larger diameter can collect a larger biopsy specimen [20]. A higher sample volume contributes to a more accurate pathological diagnosis [20]. However, in the clinical setting, a thicker needle makes the puncture procedure more difficult, particularly during transduodenal puncture in which the steep up-angle of the scope results in less flexibility and greater friction for the needle sheath. Moreover, this steep up-angle can distort the puncture needle, necessitating additional needles for a subsequent puncture. Given these facts, a needle that is flexible but can collect a sufficient volume of specimens for pathological assessment is needed for pancreatic core needle biopsy.

The EchoTip ProCore® needle (Wilson Cook Medical Inc., Winston-Salem, NC, USA) has a hollowed-out core trap on the side of the needle. This core trap system is designed to improve the acquisition of core biopsy specimens. Sterlacci et al. [21] reported that EUS tissue sampling using a 22-gauge (22G) needle with a core trap increased the accuracy of pathological diagnosis for abdominal masses compared with using a 22G aspiration needle. EchoTip ProCore® needles of various diameters are currently available. In Japan, the EchoTip ProCore 22G® (PC22) was approved for pancreatic EUS-FNA/B in March 2012 and the EchoTip ProCore 20G® needle (PC20) in December 2015. Notably, the PC20 has an anterograde, forward-bevel core trap, while the PC22 has a retrograde, reverse-bevel core trap. Additionally, the PC20 employs a new coil sheath resulting in greater flexibility of the needle, making procedures involving the duodenum easier. A recent study using porcine livers showed preferable tissue acquisition with the PC20 [22]. However, whether the PC20 can improve the diagnostic capability of EUS-FNA/B for pancreatic cancer in the real-world clinical setting has not been proven. In the present study, we retrospectively assessed the diagnostic efficacy and safety of the PC20 in patients with pancreatic cancer.

Methods

Patients

In this retrospective study, we investigated the clinical database of Kyorin University Hospital. We analyzed 191 patients who were diagnosed with pancreatic cancer and underwent their first EUS-FNA/B with the PC20, PC22, or Acquire 22G® (AC22) (Boston Scientific, Natick, MA, USA) for pancreatic solid lesions from April 2013 to October 2019. The final diagnosis of pancreatic cancer was confirmed by pathological diagnosis with EUS-FNA/B, percutaneous pancreatic biopsy, open biopsy, or ascites cytology. Some patients underwent EUS-FNA/B multiple times. Our board-certified pathologists at Kyorin University Hospital assessed both cytology and histology. The pathological diagnosis of each specimen was confirmed by multiple board-certified pathologists. A histological specimen from each pass was assessed. The diagnosis of pancreatic cancer was made when adenocarcinoma was histologically diagnosed in a specimen. In cytologic assessment, a specimen with class IV and V cells was diagnosed as pancreatic cancer. For specimens that could not be definitively diagnosed by these procedures, the diagnosis of pancreatic cancer was made based on the patient’s clinical course with scheduled imaging tests during the 12-month period following EUS-FNA/B.

Ethics statement

The present study was approved by the Institutional Review Board of Kyorin University School of Medicine (IRB No. 1099) on 27 March 2018. This study was performed in accordance with the guidelines of the Declaration of Helsinki. Informed consent was obtained using an opt-out method.

Procedure of EUS-FNA/B

All patients underwent conscious sedation by administration of midazolam (Maruishi Pharmaceutical, Osaka, Japan) and pentazocine (Maruishi Pharmaceutical). The EUS-FNA/B procedures were performed using a linear array echoendoscope (GF-UCT260®; Olympus Medical Systems, Tokyo, Japan or EG-580UT®; Fujifilm Medical, Tokyo, Japan) connected to a processor featuring a color Doppler function (EU-ME1, 2®; Olympus Medical Systems or SU-1®; Fujifilm Medical, respectively). EUS-FNA/B was performed by physicians with > 5 years of experience in EUS or by fellows in our department under the supervision of these experienced physicians. The target lesion for FNA/B was visually confirmed with EUS, and color Doppler was used to evaluate the vascularity along the puncture route. The stroke length is defined as the length of a tumor measured by EUS at the puncture line. The pancreatic solid tumor was punctured through the transgastric or transduodenal route with the PC20, PC22, or AC22. After the needle reached the target, the stylet was removed and suction was applied using a 10-ml syringe. For every puncture, 20 to-and-fro movements were employed. After removing the needle, the stylet was inserted and a sample was taken and placed on a glass slide. The filamentous sample was immersed in a 10% formalin solution for histological diagnosis, and the remaining specimen on the glass slide was smeared for cytological diagnosis. Following each pass, the endoscopists assessed the specimens by macroscopic on-site evaluation (MOSE). If a large amount of bleeding occurred at the first pass, a second pass was performed by changing the suction method to the slow-pull method. If the specimens were macroscopically insufficient, more passes were performed (maximum of six passes).

MOSE

The result of MOSE was classified into three categories. In a filamentous sample of ≥ 5 mm, white tissue was defined as an adequate specimen and a reddish sample containing a large amount of blood as a clotted specimen. A sample of < 5 mm was defined as a fragment specimen.

Accuracy of diagnosis of pancreatic cancer

Patients who were diagnosed with pancreatic cancer by the first EUS-FNA/B were defined as those who had obtained an “accurate diagnosis” by EUS-FNA/B. The accuracy of EUS-FNA/B was calculated as the proportion of patients who obtained an accurate diagnosis by EUS-FNA/B among all patients with pancreatic cancer.

Statistical analysis

The chi-squared test and Fisher’s exact test were performed to compare patient backgrounds (sex, use of antiplatelet/anticoagulant agents, comorbid diabetes mellitus, smoking, gastric surgical history, location of pancreatic tumor, and puncture line), the accuracy of the pancreatic cancer diagnosis by EUS-FNA/B, the incidence of needle distortion, and the incidence of adverse events among the different core biopsy needles. For multiple comparisons of diagnostic accuracy, Bonferroni correction was employed to calculate the adjusted p value (p-adj). The Kruskal–Wallis test and Dunn’s test were performed to compare the patients’ age, maximum diameter of the tumor measured on contrast-enhanced computed tomography images, maximum stroke length for puncture measured on EUS images, puncture stroke length, and number of passes between needles. Statistical significance was defined as p < 0.05. GraphPad Prism version 8.1.2 (GraphPad Software, San Diego, CA, USA) was used for the statistical analyses.

Results

Patients’ backgrounds

We analyzed 191 patients who underwent their first EUS-FNA/B with the PC20, PC22, or AC22 from April 2013 to October 2019 at Kyorin University Hospital and were eventually diagnosed with pancreatic cancer. In Japan, the PC20, PC22, and AC22 were approved for EUS-FNA/B for pancreatic solid lesions by the Ministry of Health, Labour and Welfare in December 2015, March 2012, and October 2016, respectively. Among the 191 patients, the first EUS-FNA/B was performed in 53 patients from April 2013 to November 2015, in 13 patients from December 2015 to September 2016, and in 125 patients from October 2016 to October 2019. The PC20, PC22, and AC22 were used for 111, 68, and 12 patients, respectively. The patients’ characteristics are shown in Table 1. There was no significant difference in sex, age, tumor location, or puncture route among the needles. However, the maximum diameter of the tumor measured on contrast-enhanced computed tomography images was significantly different among the needles (PC20, 37.9 ± 1.2 mm; PC22, 32.9 ± 1.2 mm; AC22, 22.7 ± 2.6 mm; p < 0.0001). Although a sufficient stroke length of puncture was secured in all patients, the maximum stroke length for puncture measured on EUS images was shorter with the AC22 (22.1 ± 2.2 mm) (p < 0.01 and p < 0.01 compared with PC20 and PC22, respectively), while there was no significant difference between the PC20 and PC22 (30.6 ± 0.7 and 30.3 ± 0.8 mm, respectively; p > 0.99). The number of passes was significantly higher with the PC22 (4.1 ± 0.7) than PC20 (3.7 ± 0.9, p < 0.01) and AC22 (3.5 ± 0.6, p < 0.05).

Table 1 Patients characteristics

Diagnostic accuracy of EUS-FNA/B needles for pancreatic cancer

The diagnostic accuracy for pancreatic cancer with the PC20 was 96.4% (107/111), 81.1% (90/111), and 96.4% (107/111) by histology, cytology, and the combination of histology and cytology, respectively (Table 2). The accuracy with the PC22 was 58.8% (40/68), 63.2% (43/68), and 72.1% (49/68) and that with the AC22 was 75.0% (9/12), 83.3% (10/12), and 91.7% (11/12), respectively (Table 2). The diagnostic accuracy with the PC20 by histology, cytology, and the combination of histology and cytology was significantly better than that with the PC22 (p-adj < 0.0001, p-adj < 0.05, and p-adj < 0.0001, respectively), while there was no significant difference between the PC20 and AC22 (p-adj = 0.06, p-adj > 0.99, and p-adj > 0.99, respectively). The addition of cytological assessment to histological assessment improved the diagnostic accuracy with the PC22 and AC22 but not with the PC20. Comparison of diagnostic accuracy between two needles (PC20 vs. PC22, PC20 vs. AC22 and PC22 vs. AC22) are shown in Additional file 1: Table S1. Meanwhile, the diagnostic accuracy in each pancreatic location tended to be better in PC20 and AC22 compared to PC22 (Additional file 2: Table S2). Next, we examined the diagnostic efficacy of the first pass. The histological accuracy in the first pass was 85.6 (95/111) with the PC20, 33.8% (23/68) with the PC22, and 66.7% (8/12) with the AC22 (Table 3). The accuracy was significantly higher with the PC20 than PC22 (p-adj < 0.0001), while there was no significant difference between the PC20 and AC22 (p-adj = 0.32). The proportion of adequate specimens by MOSE among the samples from the first pass was 63.9% (71/111) with the PC20, 41.2% (28/68) with the PC22, and 50.0% (6/12) with the AC22. There was no significant difference among the needles (p = 0.20) (Table 3, Additional file 3: Table S3). The histological accuracy among adequate specimens and clotted specimens was 81.7% and 92.5% with the PC20, 39.3% and 30.0% with the PC22, and 83.3% and 50.0% with the AC22.

Table 2 Diagnostic accuracy for pancreatic cancer
Table 3 Histological accuracy and adequacy in macroscopic on-site evaluation of specimens from 1st pass

Requirement of additional EUS-FNA/B needles

Among the 111 cases in which the PC20 was used, 5 (4.5%) required a biopsy needle change. The needle was distorted and the stylet could not be inserted in two of these five cases. In the other three cases, the needles were changed because of insufficient specimens (i.e., fragment specimens by MOSE). The PC22 was changed in 7 cases because of fragment specimens (10.3%, 7/68). Two cases in which the AC22 was used required a needle change because of fragment specimens (16.7%, 2/12). There was no significant difference in the requirement of a needle change among the different needle types (p = 0.20) (Table 4).

Table 4 The cases using additional needles

Adverse events related to EUS-FNA/B procedure

One patient in the PC20 group developed mild pancreatitis (0.9%, 1/111). The patient recovered with conservative treatment (fasting, fluid replacement, protease inhibitor, and antibiotic therapy). No adverse events occurred among the patients in the PC22 group (0.0%, 0/68) or AC22 group (0.0%, 0/12) (p = 0.54).

Discussion

In this retrospective study, we examined the diagnostic efficacy of three puncture needles (PC20, PC22, and AC22) for the first EUS-FNA/B in patients with pancreatic cancer. The PC20 and PC22 employ a core trap system, which contributes to increased tissue acquisition and preservation of the histological structure [23]. A core trap can reportedly ensure effective cellularity for diagnosis [21, 24]. Acquiring tissue specimens with minimal structural damage increases the histopathological quality of the samples. High-quality specimens are crucial for assessment of the subtypes of malignancy (e.g., adenocarcinoma, adenosquamous carcinoma, anaplastic carcinoma, neuroendocrine carcinoma, and metastatic carcinoma) and performance of immunohistochemistry [25, 26]. Therefore, a core trap is an advantage in histopathological diagnosis of pancreatic tumors by EUS-FNA/B. However, although both the PC20 and PC22 have a core trap, the present study demonstrated that diagnostic accuracy was significantly higher with the PC20 than PC22. Two possible factors led to the difference in diagnostic accuracy between the PC20 and PC22: (1) the difference in the needle gauge and (2) the difference in the design of the core trap. A thicker needle is reportedly more useful for collection of core tissue [27] and preservation of the histological structure of specimens [18] than is a thinner needle. Meanwhile, Armellini et al. [28] reported that the forward-bevel core trap (PC20) enables the acquisition of more tissue than the reverse-bevel core trap (PC22). This is thought to be because the forward-bevel core trap catches the tissue while the needle moves forward [28]. Additionally, the core trap is larger in the PC20 than PC22 (2.9 vs. 2.0 mm, respectively). However, there is little difference in the distance from the tip to the core trap between the PC20 and PC22 (3.8 vs. 3.9 mm, respectively). This suggests that the stroke length measured by EUS imaging does not drive the needle choice between the PC20 and PC22. Our findings suggest that the PC20 is preferable for the histological diagnosis. The total accuracy (combined histology and cytology) with the PC20 (96.4%) was equal to the histological result; i.e., there was no additional diagnostic impact of cytology. Additionally, even in samples defined as clotted specimens by MOSE, the histological accuracy of clotted specimens was 92.5% with the PC20. In one study, the histological sensitivity was 92.4% in whitish tissue and 40.8% in blood clots [29]. Although a sample obtained by a thicker needle is contaminated by more blood [30], our results suggest that clotted specimens obtained by the PC20 can still be used for histological assessment.

We also assessed the diagnostic accuracy of the AC22 in this study. The AC22 is a relatively new needle with a diagnostic accuracy of 94.0–96.7% [31,32,33]. Ishikawa et al. [34] reported that the AC22 was better than a conventional needle in acquiring cells. The AC22 is designed to hold and cut the tissue with multiple sharp structures on the tip (fork tip). This characteristic structure contributes to better sample collection even in a lesion with dense fibrotic change, different from most conventional EUS-FNA/B needles [32, 35]. A randomized trial by Bang et al. [36]. demonstrated that AC22 showed better diagnostic accuracy for pancreatic masses compared to 22-gauge conventional needles and PC22 In an animal experiment, there was no significant difference in the core tissue acquisition ability between the PC20 and AC22 [22]. Considering that the AC22 is thinner (22G) than the PC20 (20G), the fork-tip structure of the AC22 could be more advantageous for tissue collection than the core trap of the PC20. However, a recent study proposed that the side bevel, which moves to and fro in the tumor during the sampling procedure, can cut the tumor surface more effectively than the needle tip within the tumor [37]. Thus, considering the tip structure of the AC22, this needle might be more effective than other types particularly for cases in which a long puncture stroke length cannot be secured. Fujita et al. [38] compared the diagnostic accuracy between the AC22 and a conventional needle and found no difference between the two needles for > 20-mm tumors; however, there was a significant difference for < 19-mm tumors. There was no significant difference in the histological accuracy between the PC20 and AC22 in the present study (p = 0.02, p-adj = 0.06). A sufficient stroke length was secured for all patients; however, given that the puncture stroke length was significantly shorter with the AC22 than PC20, it is difficult to draw a strict conclusion regarding the efficacy of the PC20 versus AC22 from this study. However, our findings suggest that as long as a sufficient stroke length can be secured, the diagnostic accuracy of both the PC20 and AC22 is sufficiently high. Given our results that the histological accuracy tended to be higher with the PC20 than AC22 as well as the small number of patients in whom the AC22 was used, we speculate that the PC20 is suitable for cases in which histological assessment including immunohistochemistry can be critical for diagnosis, while the AC22 is suitable for small tumors in which it is difficult to secure a long stroke length.

In EUS-FNA/B, transduodenal puncture is challenging in some cases because of the steep up-angle of the scope [23, 39]. A thicker needle with less flexibility has a disadvantage in this procedure, and endoscopists need to use additional needles in some cases. In this study, however, the operability of the PC20 was not different from that of the PC22 and AC22. This finding suggests that the PC20 can be used for transduodenal puncture similarly to 22G needles. In terms of adverse events, one case of mild pancreatitis was observed in the PC20 group (0.9%), and no adverse events occurred in the PC22 and AC22 groups. No serious complications such as active bleeding or perforation occurred with any needles. A systematic review on the safety of EUS-FNA revealed a complication rate of 0.98% and mortality rate of 0.02% [40]. The needle diameter is not thought to be associated with the adverse event rate [41, 42]. Considering the results of both our study and previous studies, it seems that there is no considerable difference in safety among the PC20, PC22, and AC22.

This study had several limitations. First, because it was a single-center, retrospective study, the EUS-FNA/B skills might have varied over time and/or among endoscopists, and the needle selection by each operator might have been biased. There could be a possibility that the characteristic structure of AC22, the multiple-sharp-structure tip, impacts the needle selection. Second, the AC22 was approved for EUS-FNA/B for pancreatic solid lesions in October 2016 in Japan, and in our facility, the AC22 has been used in much fewer patients than the PC20 and PC22. It is needed to perform a prospective study with randomizing the needle selection and setting the specific criteria for the tumor size and the number of passes. Additionally, in this study, the stroke length of puncture under EUS images was not controlled between the PC20 and AC22. A study in which the stroke length is identical between the needles is necessary for direct comparison of the efficacy of these needles.

Conclusions

In conclusion, this study demonstrated that the PC20 has better diagnostic efficacy than the PC22. Although further study is needed to compare the PC20 and AC22, the high histological accuracy of the PC20 could be advantageous for lesions in which histological assessment is critical, while the AC22 could be a good option particularly for small lesions. Evidence-based needle selection could contribute to the improvement of diagnosing pancreatic cancer.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AC22:

Acquire 22G®

EUS-FNA/B:

Endoscopic ultrasound-guided fine-needle aspiration/biopsy

MOSE:

Macroscopic on-site evaluation

PC20:

EchoTip ProCore 20G®

PC22:

EchoTip ProCore 22G®

References

  1. Siegel R, Ma J, Zou Z, Jemal A. Cancer statistics, 2014. CA Cancer J Clin. 2014;64:9–29.

    Article  PubMed  Google Scholar 

  2. Gillen S, Schuster T, Zum Buschenfelde CM, Friess H, Kleeff J. Preoperative/neoadjuvant therapy in pancreatic cancer: a systematic review and meta-analysis of response and resection percentages. PLoS Med. 2010;7:1000267.

    Article  CAS  Google Scholar 

  3. Ilic M, Ilic I. Epidemiology of pancreatic cancer. World J Gastroenterol. 2016;22:9694–705.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Luo J, Xiao L, Wu C, Zheng Y, Zhao N. The incidence and survival rate of population-based pancreatic cancer patients: Shanghai Cancer Registry 2004–2009. PLoS ONE. 2013;8:e76052.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. American Cancer Society. Key statistics for pancreatic cancer. https://www.cancer.org/cancer/pancreatic-cancer/about/key-statistics.html. Accessed 17 April 2020.

  6. Saad AM, Turk T, Al-Husseini MJ, Abdel-Rahman O. Trends in pancreatic adenocarcinoma incidence and mortality in the United States in the last four decades; a SEER-based study. BMC Cancer. 2018;18:688.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Pourshams A, Sepanlou SG, Ikuta KS, Bisignano C, Safiri S, Roshandel G, et al. The global, regional, and national burden of pancreatic cancer and its attributable risk factors in 195 countries and territories, 1990–2017: a systematic analysis for the Global Burden of Disease Study 2017. Lancet Gastroenterol Hepatol. 2019;4:934–47.

    Article  Google Scholar 

  8. McGuigan A, Kelly P, Turkington RC, Jones C, Coleman HG, McCain RS. Pancreatic cancer: a review of clinical diagnosis, epidemiology, treatment and outcomes. World J Gastroenterol. 2018;24:4846–61.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Parkin DM, Boyd L, Walker LC. The fraction of cancer attributable to lifestyle and environmental factors in the UK in 2010. Br J Cancer. 2011;105(Suppl 2):S2–5.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Gowing LR, Ali RL, Allsop S, Marsden J, Turf EE, West R, et al. Global statistics on addictive behaviours: 2014 status report. Addiction. 2015;110:904–19.

    Article  PubMed  Google Scholar 

  11. Facciorusso A, Martina M, Buccino RV, Nacchiero MC, Muscatiello N. Diagnostic accuracy of fine-needle aspiration of solid pancreatic lesions guided by endoscopic ultrasound elastography. Ann Gastroenterol. 2018;31:513–8.

    PubMed  PubMed Central  Google Scholar 

  12. Wang M, Huang S, Pei R, Lin J, Yang X. Endoscopic ultrasonography guided transgastric trans-portal system fine needle aspiration for diagnosing pancreatic head and uncinate process malignancy. Ann Transl Med. 2019;7:719.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Kim J, Ryu JK, Park JM, Paik WH, Song BJ, Kim YT, et al. Clinical factors associated with accuracy of EUS-FNA for pancreatic or peripancreatic solid mass without on-site cytopathologists. J Gastroenterol Hepatol. 2014;29:887–92.

    Article  PubMed  Google Scholar 

  14. Madhoun MF, Wani SB, Rastgi A, Early D, Gaddam S, Tierney WM, et al. The diagnostic accuracy of 22-gauge and 25-gauge needles in endoscopic ultrasound-guided fine needle aspiration of solid pancreatic lesions: a meta-analysis. Endoscopy. 2013;45:86–92.

    Article  CAS  PubMed  Google Scholar 

  15. Haba S, Yamao K, Bhatia V, Mizuno N, Hara K, Hijioka S, et al. Diagnostic ability and factors affecting accuracy of endoscopic ultrasound-guided fine needle aspiration for pancreatic solid lesions: Japanese large single center experience. J Gastroenterol. 2013;48:973–81.

    Article  PubMed  Google Scholar 

  16. Siddiqui UD, Rossi F, Rosenthal LS, Padda MS, Murali-Dharan V, Aslanian HR. EUS-guided FNA of solid pancreatic masses: a prospective, randomized trial comparing 22-gauge and 25-gauge needles. Gastrointest Endosc. 2009;70:1093–7.

    Article  PubMed  Google Scholar 

  17. Yusuf TE, Ho S, Pavey DA, Michael H, Gress FG. Retrospective analysis of the utility of endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) in pancreatic masses, using a 22-gauge or 25-gauge needle system: a multicenter experience. Endoscopy. 2009;41:445–8.

    Article  CAS  PubMed  Google Scholar 

  18. Song TJ, Kim JH, Lee SS, Eum JB, Moon SH, Park DH, et al. The prospective randomized, controlled trial of endoscopic ultrasound-guided fine-needle aspiration using 22G and 19G aspiration needles for solid pancreatic or peripancreatic masses. Am J Gastroenterol. 2010;105:1739–45.

    Article  PubMed  Google Scholar 

  19. Li H, Li W, Zhou QY, Fan B. Fine needle biopsy is superior to fine needle aspiration in endoscopic ultrasound guided sampling of pancreatic masses. Medicine. 2018;97(13):e0207.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Matsubayashi H, Matsui T, Yabuuchi Y, Imai K, Tanaka M, Kakushima N, et al. Endoscopic ultrasonography guided-fine needle aspiration for the diagnosis of solid pancreaticobiliary lesions: clinical aspects to improve the diagnosis. World J Gastroenterol. 2016;22:628–40.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Sterlacci W, Sioulas AD, Veits L, Gonullu P, Schachschal G, Groth S, et al. 22-Gauge core vs 22-gauge aspiration needle for endoscopic ultrasound-guided sampling of abdominal masses. World J Gastroenterol. 2016;22:8820–30.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Mukai S, Itoi T, Katanuma A, Irisawa A. An animal experimental study to assess the core tissue acquisition ability of endoscopic ultrasound-guided histology needles. Endosc Ultrasound. 2018;7:263–9.

    Article  PubMed  Google Scholar 

  23. Conti CB, Cereatti F, Grassia R. Endoscopic ultrasound-guided sampling of solid pancreatic masses: the fine needle aspiration or fine needle biopsy dilemma. Is the best needle yet to come? World J Gastrointest Endosc. 2019;11:454–71.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Alatawi A, Beuvon F, Grabar S, Leblanc S, Chaussade S, Terris B, et al. Comparison of 22G reverse-beveled versus standard needle for endoscopic ultrasound-guided sampling of solid pancreatic lesions. United Eur Gastroenterol J. 2015;3:343–52.

    Article  Google Scholar 

  25. Varadarajulu S, Fraig M, Schmulewitz N, Roberts S, Wildi S, Hawes RH, et al. Comparison of EUS-guided 19-gauge Trucut needle biopsy with EUS-guided fine-needle aspiration. Endoscopy. 2004;36:397–401.

    Article  CAS  PubMed  Google Scholar 

  26. Levy MJ, Jondal ML, Clain J, Wiersema MJ. Preliminary experience with an EUS-guided trucut biopsy needle compared with EUS-guided FNA. Gastrointest Endosc. 2003;57:101–6.

    Article  PubMed  Google Scholar 

  27. Iwashita T, Nakai Y, Samarasena JB, Park DH, Zhang Z, Gu M, et al. High single-pass diagnostic yield of a new 25-gauge core biopsy needle for EUS-guided FNA biopsy in solid pancreatic lesions. Gastrointest Endosc. 2013;77:909–15.

    Article  PubMed  Google Scholar 

  28. Armellini E, Manfrin E, Trisolini E, Andorno S, Ballare M, Bernardoni L, et al. Histologic retrieval rate of a newly designed side-bevelled 20G needle for EUS-guided tissue acquisition of solid pancreatic lesions. United Eur Gastroenterol J. 2019;7:96–104.

    Article  Google Scholar 

  29. Iwashita T, Yasuda I, Mukai T, Doi S, Nakashima M, Uemura S, et al. Macroscopic on-site quality evaluation of biopsy specimens to improve the diagnostic accuracy during EUS-guided FNA using a 19-gauge needle for solid lesions: a single-center prospective pilot study (MOSE study). Gastrointest Endosc. 2015;81:177–85.

    Article  PubMed  Google Scholar 

  30. Yang MJ, Yim H, Hwang JC, Lee D, Kim YB, Lim SG, et al. Endoscopic ultrasound-guided sampling of solid pancreatic masses: 22-gauge aspiration versus 25-gauge biopsy needles. BMC Gastroenterol. 2015;15:122.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Bang JY, Hebert-Magee S, Navaneethan U, Hasan MK, Hawes R, Varadarajulu S. Randomized trial comparing the Franseen and Fork-tip needles for EUS-guided fine-needle biopsy sampling of solid pancreatic mass lesions. Gastrointest Endosc. 2018;87:1432–8.

    Article  PubMed  Google Scholar 

  32. Mita N, Iwashita T, Uemura S, Iwasa Y, Toda K, Mukai T, et al. Endoscopic ultrasound-guided fine needle biopsy using 22-gauge Franseen needle for the histological diagnosis of solid lesions: a multicenter prospective pilot study. Dig Dis Sci. 2020;65:1155–63.

    Article  PubMed  Google Scholar 

  33. Bang JY, Hebert-Magee S, Hasan MK, Navaneethan U, Hawes R, Varadarajulu S. Endoscopic ultrasonography-guided biopsy using a Franseen needle design: initial assessment. Dig Endosc. 2017;29:338–46.

    Article  PubMed  Google Scholar 

  34. Ishikawa T, Kawashima H, Ohno E, Tanaka H, Sakai D, Iida T, et al. Clinical impact of EUS-guided fine needle biopsy using a novel Franseen needle for histological assessment of pancreatic diseases. Can J Gastroenterol Hepatol. 2019;2019:8581743.

    Article  PubMed  PubMed Central  Google Scholar 

  35. Iwashita T, Yasuda I, Shimizu M. Endoscopic ultrasound-guided tissue acquisition: can fork and crown cut the tissue? Dig Endosc. 2019;31:146–7.

    Article  PubMed  Google Scholar 

  36. Bang JY, Krall K, Hjala N, Singh C, Tejani M, Arnoletti JP, et al. Comparing needle and methods of endoscopic ultrasound-guided fine-needle biopsy to optimize specimen quality and diagnostic accuracy for patients with pancreatic masses in a randomized trial. Clin Gastroenterol Hepatol. 2020;8:S1542-3565.

    Google Scholar 

  37. Cheng B, Zhang Y, Chen Q, Sun B, Deng Z, Shan H, et al. Analysis of fine-needle biopsy versus fine-needle aspiration in diagnosis of pancreatic and abdominal masses: a prospective, multicenter, randomized controlled trial. Clin Gastroenterol Hepatol. 2018;16:1314–21.

    Article  PubMed  Google Scholar 

  38. Fujita A, Ryozawa S, Mizuide M, Araki R, Nagata K, Tnisaka Y, et al. Does endoscopic ultrasound-guided fine needle biopsy using a Franseen needle really offer high diagnostic accuracy? A propensity-matched analysis. Endosc Int Open. 2019;7:E1327–32.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Iwashita T, Nakai Y, Mukai T, Togawa O, Matsubara S, Hatano Y, et al. A 19-gauge histology needle versus a 19-gauge standard needle in endoscopic ultrasound-guided fine-needle aspiration for solid lesions: a multicenter randomized comparison study (GREATER Study). Dig Dis Sci. 2018;63:1043–51.

    Article  PubMed  Google Scholar 

  40. Wang KX, Ben QW, Jin ZD, Du YQ, Zou DW, Liao Z, et al. Assessment of morbidity and mortality associated with EUS-guided FNA: a systematic review. Gastrointest Endosc. 2011;73:283–90.

    Article  PubMed  Google Scholar 

  41. Varadarajulu S, Ginnetti L, Peetermans J, Rousseau M, Hasan M, Hawes R. Meta-analysis comparing ratesof complications betweenthe standard19Gand22G/25G needlesfor EUS-guidedFNAof pancreatic lesions. Gastrointest Endosc. 2013;77:AB405.

    Google Scholar 

  42. Polkowski M, Jenssen C, Kaye P, Carrara S, Deprez P, Gines A, et al. Technical aspects of endoscopic ultrasound (EUS)-guided sampling in gastroenterology: European Society of Gastrointestinal Endoscopy (ESGE) Technical Guideline- March 2017. Endoscopy. 2017;49:989–1006.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We thank Angela Morben, DVM, ELS, from Edanz Group (https://en-author-services.edanzgroup.com/), for editing a draft of this manuscript.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

SW, JM, and TH designed and conceived this study. SW, JM, and TH analyzed and interpreted the results and prepared the manuscript. SW, MT, KK, SK, TN, TG, HO, KO, KG, NI, ST, KN, and JS collected and analyzed the data. JM and TH oversaw the entire project. TH approved the final draft submitted. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Jun Miyoshi or Tadakazu Hisamatsu.

Ethics declarations

Ethics approval and content to participate

The present study was approved by the Institutional Review Board of Kyorin University School of Medicine (IRB No. 1099) on 27 March 2018. This study was performed in accordance with the guidelines of the Declaration of Helsinki. The IRB approved the use of an opt-out method of consent and informed consent was obtained using an opt-out method with the written document on the research.

Consent for publication

By opt-out of the research, we disclose the information on the bulletin board or website of the facility, guarantee the opportunity for the research subject to refuse to participate in the research, and provide a contact point (contact address) that accepts the intention to reject the research.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Table S1. Comparison of diagnostic accuracy between two needles.

Additional file 2.

Table S2. Comparison of diagnostic accuracy in each pancreatic location between two needles.

Additional file 3.

Table S3. Comparison of macroscopic on-site evaluation between two needles.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Watanabe, S., Miyoshi, J., Toki, M. et al. Effectiveness of introducing a 20-gauge core biopsy needle with a core trap in EUS-FNA/B for diagnosing pancreatic cancer. BMC Gastroenterol 21, 8 (2021). https://doi.org/10.1186/s12876-020-01583-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12876-020-01583-7

Keywords