Table 1 Reliable liver stiffness measurement

From: Possible use of 2D shear wave liver elastography in new-onset ascites evaluation

Reliable liver stiffness measurement using Esaote 2D-SWE

6 successful measurements

> 60% success of all the measurements taken

IQR/LSM < 30%

ROI > 10 mm in all measurements

SD <3x of LSM in all measurements

Depth of the measurement < 50 mm

Measurement > 1 cm below liver capsule and in a region without major vessels, bile ducts or masses

Measurement in two different intercostal spaces

Evenly distributed FOV colour

Evenly spaced colour in ROI (no blue-red jumps)

High quality measurement according to the manufacturer’s instrucitonsf

Diameter of perihepatic fluid between probe and liver < 20 mm

  1. aIQR interquartile range, bLSM liver stiffnes measurement, cROI region of interest, dFOV field of vision, eSD standard deviation. fsee Fig. 1