Skip to main content

Table 2 Pulse oxygen saturation and emergent management

From: Comparison of the innovative endoscopic oropharyngeal airway and the conventional mouthpiece in elderly outpatients undergoing esophagogastroduodenoscopy under sedation: a prospective and randomized study

 

Mouthpiece group (n = 180)

Airway group (n = 180)

P-value

Pulse oxygen saturation

   

 Subclinical respiratory depression,pulse oxygen saturation of 90–94% (%)

19 (10.6)

9 (5.0)

0.049

 Mild hyoxemia, pulse oxygen saturation of 85–89% (%)

22 (12.2)

8 (4.4)

0.008

 Moderate hyoxemia, pulse oxygen saturation < 85% (%)

2 (1.1)

0 (0.0)

0.156

Emergency management

   

 Lifting the mandible (%)

25 (13.9)

11 (6.1)

0.014

 Enhancing oxygen flow (%)

43 (23.9)

17 (9.4)

0.000

 Withholding endoscopy (%)

17 (9.4)

6 (3.3)

0.018

 Effective suction secretions and reflux (%)

5 (2.8)

15 (8.3)

0.021

 Removing the endoscopy tube and mask ventilation (%)

2 (1.1)

0 (0.0)

0.156

  1. Values are presented as n (%)