Skip to main content

Table 1 Changes of flow resistance in the bile flow phantom model

From: Evaluation of valve function in antireflux biliary metal stents

Type of antireflux valve

Flow rate (L/min)

Flow resistance

(mbar, mean ± standard deviation)

p-value

Baseline

12 weeks

Wine glass type

(n = 3)

0.25

103.8 ± 2.9

105.8 ± 2.3

0.109

0.5

154.0 ± 4.5

166.4 ± 9.5

0.109

0.75

219.8 ± 5.3

230.8 ± 17.3

0.180

1.0

282.6 ± 10.3

305.5 ± 23.4

0.109

1.25

353.6 ± 8.6

400.0 ± 44.3

0.285

Windsock type (n = 5)

0.25

36.4 ± 0.9

42.8 ± 0.9

0.039

0.5

53.6 ± 1.9

58.0 ± 2.9

0.042

0.75

69.0 ± 1.7

71.6 ± 2.8

0.223

1.0

87.6 ± 0.5

87.6 ± 1.7

1.000

1.25

105.0 ± 0.7

104.8 ± 1.6

0.854

Funnel type

(n = 5)

0.25

36.2 ± 1.3

43.2 ± 2.0

0.039

0.5

51.8 ± 0.8

59.2 ± 2.1

0.042

0.75

64.6 ± 1.1

74.4 ± 2.9

0.042

1.0

84.4 ± 1.8

89.4 ± 2.9

0.078

1.25

101.8 ± 1.1

105.6 ± 1.7

0.042

  1. Data were analyzed by using the Wilcoxon signed rank test