| EVOTECH ECR | Manual cleaning plus AER* |
---|---|---|
Phase 1 - Preparation of scopes for reprocessing* | Â | Â |
Fill sink with water | x | x |
Clean valves Remove suction, air, water and biopsy valves in sink; brush the valves and other removable parts; place valve parts in the reprocessor. | x | x |
Manual wet leak test** | x | x |
Add enzymatic detergent to sink | Â | x |
Brush internal channels three times each | Â | x |
Connect cleaning adapters/tubing to scope | Â | x |
Flush interior of scope with enzymatic detergent using hospital-developed automated flushing system | Â | x |
Drain sink of enzymatic detergent | Â | x |
Rinse interior of scope with tap water using hospital-developed automated flushing system | Â | x |
Rinse exterior of scope with tap water spray | Â | x |
Drain sink and remove scope from sink | x | x |
Dry exterior of scope with a blue wipe | Â | x |
Walk from sink to reprocessor with scope | x | x |
Place scope in an unoccupied basin of reprocessor | x | x |
Connect colour coded connecting tubing to scope according to specific scope connection diagram | x | Â |
Enter identification data for operator, physician, scope type, procedure and patient*** | x | Â |
Connect reprocessing adapter/tubing to connection port in basin | Â | x |
Place lid closure on reprocessor | x | x |
Program and start cycle | x | x |
Phase 2 - Removal of scopes from reprocessor at conclusion of cycle # | Â | Â |
Enter operator identification | x | x |
Open reprocessor lid | x | x |
Disconnect color coded tubing from scope | x | Â |
Disconnect reprocessing adapter/tubing from connection port in basin with manual check to ensure connections were properly placed | Â | x |
Return accessories and scope to Cleanascope® transport tray for next procedure or to the storage cupboard if the scope is not required in the short term | x | x |