Skip to main content
  • Research article
  • Open access
  • Published:

Does the variable-stiffness colonoscope makes colonoscopy easier? A meta-analysis of the efficacy of the variable stiffness colonoscope compared with the standard adult colonoscope

Abstract

Background

The variable-stiffness colonoscope (VSC) appears to have advantages over the standard adult colonoscope (SAC), although data are conflicting. To provide a comprehensive up-to-date review, we conducted a meta-analysis to compare the efficacies of the VSC and SAC.

Methods

Electronic databases, including PubMed, EMBASE, the Cochrane library and the Science Citation Index, were searched to retrieve relevant trials. In addition, meeting abstracts and the reference lists of retrieved articles were reviewed for further relevant studies.

Results

Eight randomized controlled trials (RCTs), enrolling a total of 2033 patients, were included in the meta-analysis. There was no significant heterogeneity among these studies. The cecal intubation rate was higher with the use of VSC (RR = 1.03, 95% CI 1.01 to 1.06, 8 RCTs). The VSC was also associated with fewer position changes made during colonoscopy. Time to cecal intubation was similar with VSC and SAC (WMD −0.54, 95% CI −1.40 to 0.32) but shorter in subgroup analysis with the use of VSC (WMD = −1.36, 95% CI −2.29 to −0.43). Sedation dose used with the two types of instruments showed no evidence of differences either. For all trials, only patients were blinded because of the nature of the interventions.

Conclusion

Use of the VSC significantly improved the cecal intubation rate and reduced ancillary maneuvers made during the procedure. Cecal intubation time was similar for the two colonoscope types over all trials, whereas a shortened time with the use of the adult VSC was seen in subgroup analysis.

Peer Review reports

Background

With the increasing availability of colonoscopy, it has become the most common and accurate tool for detecting structural lesions of the lower gastrointestinal tract and for diagnosing colonic diseases, such as colorectal cancer, polyps and inflammatory bowel disease. However, the presence of sharp angulations or looping always increases the difficulty of the procedure and causes patients distinct discomfort. The failure rate to achieve the cecum initially remains significant at up to 2%-10% [13]. Although colonoscopic technical proficiency has demonstrated widespread diffusion throughout the developed world, modifications or improvements should to made to ameliorate outcomes.

The variable-stiffness colonoscope (VSC), which can be incorporated into the standard adult and pediatric colonoscope chasses, has a stiffness control ring with dial setting that ranges from 0 to 3. The endoscopist can adjust the relative flexibility of the scope’s insertion tube. VSC is now available in ‘adult’ and ‘pediatric’ sizes. It has been suggested that this type of colonoscope has a theoretical advantage over the standard adult colonoscope (SAC) with its smaller diameter and greater flexibility. In recent years, several studies have been performed to compare VSC with SAC. However, the results have been inconclusive and could not determine whether VSC is superior to SAC or more suitable for routine adult colonoscopy [49]. Therefore, we conducted a meta-analysis to compare the efficacy of VSC with SAC.

Methods

Literature search

First, electronic databases, including PubMed (1966 to November 2011), EMBASE (1980 to November 2011), the Cochrane Central Register of Controlled Trials (CENTRAL, The Cochrane Library, Issue 4 of 4, Oct 2011), and the Science Citation Index were searched. The search strategy was performed with the following search terms as both free-text terms as well as MeSH terms: colonoscope, colonoscopy, stiff*, stiffness, variable stiffness colonoscopy, pediatric variable stiffness colonoscopy. Second, meeting abstracts and the reference lists of retrieved articles were reviewed for additional relevant studies. No language restriction was imposed.

Study selection

Randomized controlled trials (RCTs) comparing VSC with SAC were included for analysis. Only the most recent study was included if more than one study was published using the same study population. Open, uncontrolled, observational studies and case reports were excluded from the meta-analysis.

Data abstraction

All the data were tabulated with standard data abstractions sheets. For each study and each type of intervention, the following characteristics were extracted: study design and conduct, numbers of patients, endoscopist characteristics, instrument features and study outcomes. Study outcomes included cecal intubation rate, cecal intubation time, sedation dose used, abdominal pain score, and ancillary maneuvers during the procedure (manual pressure used and position changes made).

Two investigators (Xie Q, Chen B) independently extracted details of study population, interventions and outcomes. The paper was reviewed if either one of the two investigators thought an abstract was relevant. If there were any discrepancies about information given in the title and abstract, the full article was reviewed for clarification. Differences in opinion were resolved by discussing with the third author (Liu L).

Assessment of risk of bias in included studies

For the risk of bias assessment, two investigators independently used an assessment form recommended by the Cochrane Handbook. Any disagreements were resolved by a third author until consensus was obtained. We considered the following criteria:

  1. 1.

    Sequence generation: Was the allocation sequence adequately generated?

  2. 2.

    Allocation concealment: Was the allocation adequately concealed?

  3. 3.

    Blinding: Was knowledge of the allocated intervention adequately prevented during the study?

  4. 4.

    Incomplete outcome data: Were incomplete outcome data adequately addressed?

  5. 5.

    Selective outcome reporting: Were reports of the study free of suggestion of selective outcome reporting?

  6. 6.

    Other sources of bias: Was the study apparently free of other problems that could put it at a high risk of bias?

Each domain was graded as yes (low risk of bias), no (high risk of bias), or unclear (uncertain risk of bias) according to the criteria.

For ranking the strength and quality of the evidence for a given comparison, the GRADE and Summary of Findings tables recommended by the Cochrane Collaboration were used.

Assessment of reporting biases

For the assessment of publication bias, a funnel plot was conducted if sufficient data were available.

Statistical analysis

Meta-analyses were conducted for trials comparing VSC with SAC, using the statistical tool Revman 5.1. Dichotomous data were expressed as relative risk (RR) or odds ratio (OR) and continuous outcomes as the weighted mean difference (WMD) with 95% confidence interval (CI). A fixed effects model was used for pooling of data when statistical heterogeneity was not present. If heterogeneity was existed, a random effects model was performed.

Heterogeneity was quantified with Cochran’s Q test and the I2 metric, and 95% CI for I2 were calculated. I2 was in a scale of 0-100%. If there was “considerable heterogeneity”, which is defined by the Cochrane Handbook for Systematic Reviews of Interventions as an I2 value between 75% and 100%, the data were not pooled. When I2 >50%, suggesting very large heterogeneity between studies, the random effects model was used and a sensitivity analysis was planned to evaluate heterogeneity among studies.

Results

Search results

Overall, 32 articles were identified comparing VSC with SAC. After reading abstracts and full-texts, we excluded 24 of these articles [427], because they were reviews or not RCTs. Finally, 8 studies met the criteria for inclusion in the review [2835].

Trial characteristics

The characteristics of these studies are summarized in Table 1. All these studies were RCTs, containing a total of 2033 participants (1041 male, 992 female), aged from 15 to 89 years. Previous abdominal or pelvic surgery was reported for 13% of the participants in seven trials [28, 3035]. Seven trials listed procedure indications [28, 3035], with screening colonoscopy or polyp surveillance being the main indications. Five trials compared adult VSC with SAC [2830, 34, 35], while the other three were performed with pediatric VSCs [3133].

Table 1 The characteristics of included trials comparing the VSC with the SAC

The instruments used in the trials included VSCs, pediatric VSCs and SACs. The adult VSCs used were: the Olympus XCF-SH140I (distal tip diameter of 13.2 mm, shaft diameter of 12.9 mm, instrument channel of 3.2 mm, working length 133 cm); the Olympus XCF-SH230L (shaft diameter 12.9 mm, instrument channel 3.2 mm, working length 168 cm); the Olympus CF-240AI (shaft diameter 12.0 mm, instrument channel of 3.2 mm, working length 138 cm); the Olympus CF-Q1402 (shaft diameter of 12.0 mm, instrument channel of 3.2 mm, working length 168 cm) and the Olympus CF-H180AI/L (shaft diameter 12.8 mm, working length of 168 cm). The pediatric VSCs used were: the Olympus XPCF-140AL (insertion tube outer diameter 11.3 mm, instrument channel of 3.2 mm, working length 168 cm); the Olympus PCF-Q260AI (insertion tube outer diameter 11 mm, instrument channel of 3.2 mm, working length of 138 cm) and the Olympus PCF-160AL (insertion tube outer diameter of 11.5 mm, instrument channel of 3.2 mm, working length 168 cm).

The level of experience of endoscopists was evaluated either by years of experience (7 to 20 years) or by the number of procedures done (more than 500 procedures). In the retrieved articles, seven trials evaluated VSC procedures with experienced endoscopists; while two studies evaluated VSC procedures among less experienced endoscopists (one study included both experienced and less experienced endoscopists).

Risk of bias in included studies

Among the eight RCTs included in this meta-analysis, the allocation sequence was generated using a random number generator [28]; computer-generated random number table [34]; and pre-randomized cards [29]. Four of the eight trials reported adequate allocation concealment [29, 30, 33, 35], while in another four trials the allocation concealment was unclear. The eight trials blinded all patients but none of these trials blinded the endoscopists because of the nature of interventions.

The quality of the evidence for the outcomes for the included studies is shown in the Summary of Findings table (Table 2)

Table 2 Summary of findings for the main comparison

Outcomes

Cecal intubation rate

Eight trials assessed VSC or pediatric VSC vs. SAC, and the cecal intubation rate was higher in VSC compared with SAC (RR 1.03, 95% CI 1.01 to 1.06, Figure 1). There was no heterogeneity among these studies (I2 = 0%, P = 0.59).

Figure 1
figure 1

Cecal intubation rate comparing VSC with SAC; relative risk (RR) with 95% confidence intervals (CI). Yoshikawa 2002(1) represented the experienced group; Yoshikawa 2002(2) represented the limited experience group. VSC: variable-stiffness colonoscope; SAC: standard adult colonoscope.

Cecal intubation time

Meta-analysis of seven trials showed no significant difference of cecal intubation time between SAC and VSC (mean difference −0.54, 95% CI −1.40 to 0.32; Figure 2). There was medium heterogeneity among these seven studies (I2 = 47%, P = 0.07).

Figure 2
figure 2

Cecal intubation time comparing VSC with SAC; mean differences with 95% CI.

Sedation dosage

Seven studies reported sedation dose used during the procedure but only five studies listed these data as mean and standard deviation. One trial used a patient-controlled analgesia (PCA) pump consisting of a mixture of propofol and alfentanil [34]; another one employed a combination of midazolam, propofol and fentanyl [35]. Shumaker et al. reported a mixture use of meperidine, midazolam and fentanyl [30], and the last two studies used midazolam and meperidine together [29, 33]. Only the doses of meperidine and midazolam intravenously administered were calculated in view of the different types of data recorded. The doses of meperidine (WMD = 1.32, 95%CI −3.64 to 1.01, three trials) and midazolam (WMD = −0.03, 95% CI −0.15 to 0.08, four trials) were similar with the use of either VSC or SAC.

Abdominal pain

Five studies presented pain scores as mean and standard deviation. However, the scales used for scoring pain were different. In two studies [30, 34], a 0 to 10 score scale was used, and the other three used 0 to 9, 0 to 4 and 0 to 100 visual analogue scales, respectively [29, 31, 35]. Due to the differences in the scale, we did not pool the data for these studies.

Ancillary maneuvers

Seven trials listed the data on abdominal pressure use during the procedure. The odds for the use of abdominal pressure during the procedure were similar in both groups (RR 0.92, 95% CI 0.75 to 1.12, Figure 3).

Figure 3
figure 3

Use of abdominal pressure during colonoscopy with VSC and SAC; relative risk (RR) with 95% CI.

VSC was associated with fewer position changes made during colonoscopy (OR 0.65, 95% CI 0.47 to 0.89). The meta-analysis among the five studies showed low heterogeneity (I2 = 44%, P = 0.13, Figure 4).

Figure 4
figure 4

Position changes made during colonoscopy with VSC and SAC; odds ratios (OR) with 95% CI.

Subgroup and sensitivity analysis

To assess the effect of heterogeneity, subgroup and sensitivity analysis were conducted.

Subgroup analysis was done to evaluate the cecal intubation rate and time during colonoscopy, according to the type of VSC (adult or pediatric VSC). The cecal intubation rate with pediatric VSC was similar to that of SAC (OR 1.57, 95% CI 0.85 to 2.90, three trials), while the odds for achieving cecal intubation were a little higher with adult VSC than SAC (RR = 1.04, 95% CI 1.01 TO 1.07, six trials, Figure 5). The cecal intubation time was similar between pediatric VSC and SAC (WMD = 0.19, 95% CI −1.04 to 1.41, three trials.), while comparing adult VSC with SAC, the cecal intubation time was relatively shorter during the procedures for adult VSC (WMD = −1.36, 95% CI −2.29 to −0.43, five trials, Figure 6).

Figure 5
figure 5

Cecal intubation rate: subgroup analysis of trials comparing adult VSC with SAC; relative risk (RR) with 95% CI.

Figure 6
figure 6

Cecal intubation time: subgroup analysis of trials comparing adult VSC with SAC; mean differences with 95% CI.

Sensitivity analysis was performed to detect the effect of any one of the included trials on the overall estimate by excluding one of them according to sample size. There were no significant changes to ORs or RRs and WMDs when excluding any one of the included trials.

To detect publication bias, asymmetry was explored in a funnel plot. Distribution of the results of each study in a funnel plot ruled out any potential publication bias.

Discussion

This meta-analysis included eight RCTs published up to November 2011, including a total of 2033 participants who received VSC or SAC. VSC showed improved cecal intubation rate, as well as decreased position change, but no evidence of advantages in cecal intubation time, sedation dose and manual pressure used. Outcomes were also analyzed in two subgroups based on the type of instruments (pediatric VSC and adult VSC) to evaluate cecal intubation rate and time. By contrast, adult VSC shortened the time to achieve cecal intubation compared with SAC, which was the only difference compared with the results considering all groups together.

In the comparison of VSC with SAC, none of the individual studies had shown an advantage in terms of the frequency of cecal intubation, but the pooled data slightly favored VSC. The increased sample size could be the most significant reason explaining the difference in cecal intubation rates. Large numbers of participants reduced the sampling error, which influenced the significance of the difference of the cecal intubation rates between VSC and SAC. This result is meaningful in clinical practice since, as we know, the failure rate for cecal intubation remains high with the use of SAC, so that this part of the anatomy does not receive clear and early diagnosis and treatment. VSC increased the intubation rate, which contributes to an early and accurate diagnosis.

The individual studies included in this meta-analysis had yielded somewhat conflicting data on cecal intubation time. Five trials found no significant difference in the time to cecal intubation, whereas the other 3 reported significantly shortened time to reach the cecum with VSC. The pooled results with all trials showed no significance; however, subgroup analysis reported shortened time with adult VSC.

Compared with the meta-analysis performed by M. O. Othman et al. [11], our results showed differences in sedation dose. In the subgroup analysis of our meta-analysis, the sedation dose used during colonoscopy was similar between VSC and SAC, but had been reduced with the use of VSC in the previous analysis [11]. In addition, we did not pool the data of pain scores for patients due to the differences in the scale.

Several other individual publications have reported discrepant results with the use of VSC and SAC. For example, Odori et al. [6] reported a prospective RCT of two prototypes of instruments in 352 consecutive cases and found that the cecal intubation time was significantly shorter with the use of VSC. VSC also reduced the need for abdominal pressure and position changes. Rex et al. [8] evaluated the cecal intubation time in a cohort study of 358 consecutive sedated participants amongst VSC, pediatric VSC and SAC. No significant difference was found in the time to reach the cecum. Kaffes et al. [9] found no evidence of difference in cecal intubation time between the two colonoscopic instruments in a nonrandomized trial of 803 participants.

In addition, different methods of activating the variable-stiffness function were used in the included studies and might explain part of the variability. In 2003, Ginsberg [4] described a ‘standard’ technique for using the VSC: the colonoscopy is started with the ‘minimum’ or ‘soft’ control ring (dial setting 0) until looping occurs or the sigmoid colon is traversed. Then, the users reduce the loop, straighten the colonoscope and increase the stiffness control to the ‘hard’ position (dial setting 3). If a loop forms again, the stiffness dial is turned to the ‘soft’ position and the process above is repeated. Horiuchi et al. [32], Sorbi et al. [29] and Sola-Vera J et al. [35] used the same approach described by Ginsberg. Yoshikawa et al. [31] adopted a similar technique to Ginsberg’s, and applied abdominal pressure when activation of the maximum setting (dial setting 3) failed to advance the colonoscope. Al-Shurieki et al. [33] made a slight modification, using the stiffening feature intermittently. When significant looping occurred, dial setting 2 was employed initially,and then stiffening to dial setting 3 was applied if the second setting failed. Lee et al. [34] began at default dial setting 0 and activated the stiffer modes (dial settings 2 or 3) if looping was encountered. Shumaker et al. [30] activated the maximum stiffness mode when the colonoscope was inserted to 30 cm from the anus and left the stiffness setting to full ‘on’ position until the cecum was achieved. The stiffness mechanism was deactivated during withdrawal. In almost all the studies reviewed above, the variable stiffness function was activated when looping was encountered. Furthermore, Shah et al. [7] performed an RCT to evaluate the effect of routinely stiffening the straightened VSC after traversing the sigmoid colon, finding that with the stiffening function activated, the time needed to negotiate the proximal colon and splenic flexure shortened and ancillary maneuvers were reduced. These results may reflect another way in which to use the VSC in clinical practice and may explain the differences in cecal intubation time in comparison with SAC.

There were no scope-related complications reported in the studies included in this meta-analysis. To this point, no safety concerns have been raised with the use of VSCs. However, a single case report draws a possible association between the use of pediatric VSC and a sigmoid perforation, and only the distal descending colon was reached in a patient with a fixed and angulated sigmoid colon [5]. During the procedure, precise judgment and caution must be used, especially when advancing through a narrowed colon or pushing through loops.

A potential limitation of this meta-analysis is that these studies could not be performed to ‘blind’ the endoscopists to the nature of the interventions. Additionally, different models and manufacturers of VSC were used in the studies included. Furthermore, indications for activating the variable stiffness function did not follow the same criteria. There was no universal method for using VSC across the studies and across large tertiary centers, which may limit generalization to other practice settings. Finally, in several studies, specific patient subsets, such as colonic cancer and prior colonic surgery, were excluded.

Conclusions

In conclusion, the present meta-analysis demonstrated that the VSC was associated with a higher likelihood of achieving cecal intubation and with fewer position changes. However, for most patients, they don’t matter which instrument was used. The more important problem appears to be how to translate these results of colonoscope trials into clinical practice. It might be difficult to predict beforehand which patients will have fixed, angulated sigmoid colons or long, floppy colons. Therefore, there may be no optimal colonoscope model for all patients and endoscopists at all times [4, 10] and further studies should be performed to confirm the role of VSC.

Author's contributions

QX participated in the design of the study, performed the statistical analysis and drafted the manuscript. BC participated in the statistical analysis and helped to draft the manuscript. LL carried out the statistical analysis. HTG participated in the design of the study. All authors read and approved the final manuscript.

References

  1. Church JM: Complete colonoscopy: how often? And if not, why not?. Am J Gastroenterol. 1994, 89: 556-560.

    CAS  PubMed  Google Scholar 

  2. Mitchell RM, McCallion K, Gardiner KR, Watson RG, Collins JS: Successful colonoscopy; completion rates and reasons for incompletion. Ulster Med J. 2002, 71: 34-37.

    CAS  PubMed  PubMed Central  Google Scholar 

  3. Cardin F, Minicuci N, Andreotti A, Pinetti E, Campigotto F, Don BM, Martella B, Terranova O: Mximizing the general success of cecal intubation during propofol sedation in a multi-endoscopist academic center. BMC Gastroenterol. 2010, 10: 123-10.1186/1471-230X-10-123.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Ginsberg GG: Colonoscopy with the variable stiffness colonoscope. Gastrointest Endosc. 2003, 58: 579-584. 10.1067/S0016-5107(03)01873-X.

    Article  PubMed  Google Scholar 

  5. Rodriguez SA, Ormseth E, Tsuchida A: Bowel perforation with the variable stiffness colonoscope. Gastrointest Endosc. 2003, 57: 271-273. 10.1067/mge.2003.58.

    Article  PubMed  Google Scholar 

  6. Odori T, Goto H, Arisawa T, Niwa Y, Ohimiya N, Hayakawa T: Clinical results and development of variable-stiffness video colonoscopes. Endoscopy. 2001, 33: 65-69. 10.1055/s-2001-11174.

    Article  CAS  PubMed  Google Scholar 

  7. Shah SG, Brooker JC, Williams CB, Thaper C, Suzuki N, Saunders BP: The variable stiffness colonoscope:assessment of efficacy by magnetic endoscope imaging. Gastrointest Endosc. 2002, 56: 195-210.

    Article  PubMed  Google Scholar 

  8. Rex DK: Effect of variable stiffness colonoscopes on cecal intubation times for routine colonoscopy by an experienced examiner in sedated patients. Endoscopy. 2001, 33: 60-64. 10.1055/s-2001-11179.

    Article  CAS  PubMed  Google Scholar 

  9. Kaffes AJ, Mishra A, Ding SL, Hope R, Williams SJ, Gillespie PE: Bourke MJ.A prospective trial of variable stiffness pediatric vs. standard instrument colonoscopy. Gastrointest Endosc. 2003, 58: 685-689. 10.1016/S0016-5107(03)02017-0.

    Article  PubMed  Google Scholar 

  10. Baron TH: The variable stiffness colonoscope:A scope for all reasons?. Am J Gastroenterol. 2002, 97: 2942-2943. 10.1111/j.1572-0241.2002.07091.x.

    Article  PubMed  Google Scholar 

  11. Othman MO, Bradley AG, Choudhary A, Hoffman RM, Roy PK: Variable stiffness colonoscope versus regular adult colonoscope: meta-analysis of randomized controlled trials. Endoscopy. 2009, 41: 17-24. 10.1055/s-0028-1103488.

    Article  CAS  PubMed  Google Scholar 

  12. Martinez JL, Calleja JL: A new variable stiffness colonoscope makes colonoscopy easier: a randomized controlled trial. Gastrointest Endosc. 2001, 53 (6): 694-695.

    CAS  PubMed  Google Scholar 

  13. Saunders BP, East JE: What are the benefits of the variable stiffness colonoscope?. Nat Clin Pract Gastroenterol Hepatol. 2008, 5 (1): 8-9.

    Article  PubMed  Google Scholar 

  14. Saunder BP, Williams CB: The variable-stiffness colonoscope:”too stiff or not too stiff, that is the question:– a new twist to the plot. Endoscopy. 2001, 33 (1): 70-73. 10.1055/s-2001-11180.

    Article  Google Scholar 

  15. Bell GD, Burn K: Measurements of the stiffness of endoscopes –a plea for commonality. Gut. 2001, 49: 154-158. 10.1136/gut.49.1.154.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Subramanian S, Rex DK: Variable stiffness colonoscopes: do they offer a better examination?. Curr Opin Gastroenterol. 2003, 19 (5): 492-496. 10.1097/00001574-200309000-00011.

    Article  PubMed  Google Scholar 

  17. Shah SG, Saunders BP: Aids to insertion:magnetic imaging, variable stiffness, and overtubes. Gastrointest Endosc Clin N Am. 2005, 15 (4): 673-686. 10.1016/j.giec.2005.08.011.

    Article  PubMed  Google Scholar 

  18. Leung FW: Methods of reducing discomfort during colonoscopy. Dig Dis Sci. 2008, 53 (6): 1462-1467. 10.1007/s10620-007-0025-9.

    Article  PubMed  Google Scholar 

  19. Baumann UA: Does retroflexion improve adenoma miss rates on the right side of the colon when using a pediatric variable-stiffness colonoscope during routine colonoscopy. Endoscopy. 2009, 41 (7): 654-10.1055/s-0029-1214876.

    Article  CAS  PubMed  Google Scholar 

  20. Chen PJ, Shih YL, Chu HC, Chang WK, Hsieh TY, Chao YC: A prospective trial of variable stiffness colonoscopes with different tip diameters in unsedated patients. Am J Gastroenterol. 2008, 103 (6): 1365-1371. 10.1111/j.1572-0241.2008.01812.x.

    Article  PubMed  Google Scholar 

  21. Hsieh YH, Zhou AL, Lin HJ: Comparing different methods of activating the variable stiffness function of a pediatric variable stiffness colonoscope. J Clin Med Assoc. 2008, 71 (1): 23-29. 10.1016/S1726-4901(08)70068-6.

    Article  Google Scholar 

  22. Hawari R, Pasricha PJ: Going for the loop:a unique overtube for the difficult colonoscopy. J Clin Gastroenterol. 2007, 41 (2): 138-140. 10.1097/01.mcg.0000225595.01201.d2.

    Article  PubMed  Google Scholar 

  23. Horiuchi A, Nakayama Y: Colonoscopy in the sitting position:lessons learned from self-colonoscopy by using a small-caliber, variable-stiffness colonoscope. Gastrointest Endosc. 2006, 63 (1): 119-120. 10.1016/j.gie.2005.10.014.

    Article  PubMed  Google Scholar 

  24. Brown GJ, Saunders BP: Advances in colonic imaging:technical improvements in colonoscopy. Eur J Gastroenterol Hepatol. 2005, 17 (8): 785-792. 10.1097/00042737-200508000-00002.

    Article  PubMed  Google Scholar 

  25. Shumaker DA, Zaman A, Katon RM: Use of variable-stiffness colonoscope allows completion of colonoscopy after failure with the standard adult colonoscope. Endoscopy. 2002, 34 (9): 711-714. 10.1055/s-2002-33442.

    Article  CAS  PubMed  Google Scholar 

  26. Heigh RI, et al: Use of electromagnetic colonoscope to assess maneuvers associated with cecal intubation. BMC Gastroenterol. 2009, 9: 24-10.1186/1471-230X-9-24.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Seitz U, Seewald S, Bohnacker S, Soehendra N: Advances in interventional gastrointestinal endoscopy in colon and rectum. Int J Colorectal Dis. 2003, 18 (1): 12-18. 10.1007/s00384-002-0406-7.

    Article  PubMed  Google Scholar 

  28. Brooker JC, Saunders BP, Shah SG, Williams CB: A new variable stiffness colonoscope makes colonoscopy easier: a randomised controlled trial. Gut. 2000, 46: 801-805. 10.1136/gut.46.6.801.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Sorbi D, Schleck CD, Zinsmeister AR, Gostout CJ: Clinical application of a new colonoscope with variable insertion tube rigidity: a pilot study. Gastrointest Endosc. 2001, 53: 638-642. 10.1067/mge.2001.114056.

    Article  CAS  PubMed  Google Scholar 

  30. Shumaker DA, Zaman A, Katon RM: A randomized controlled trial in a training institution comparing a pediatric variable stiffness colonoscope, a pediatric colonoscope, and an adult colonoscope. Gastrointest Endosc. 2002, 55: 112-179.

    Article  Google Scholar 

  31. Yoshikawa I, Honda H, Nagata K, Kanda K, Yamasaki T, Kume K, Tabaru A, Otsuki M: Variable stiffness colonoscopes are associated with less pain during colonoscopy in unsedated patients. Am J Gastroenterol. 2002, 97: 3052-3055. 10.1111/j.1572-0241.2002.07100.x.

    Article  PubMed  Google Scholar 

  32. Horiuchi A, Nakayama Y, Kajiyama M, Fujii H, Tanaka N: Usefulness of a small-caliber, variable-stiffness colonoscope as a backup in patients with difficult or incomplete colonoscopy. American J Gastroenterology. 2004, 99: 1936-1940. 10.1111/j.1572-0241.2004.40201.x.

    Article  Google Scholar 

  33. Al−Shurieki SH, Marshall JB: Is the variable-stiffness paediatric colonoscopemore effective than a standard adult colonoscope for outpatient adult colonoscopy? A randomised controlled trial. Dig Liver Dis. 2005, 37: 698-704. 10.1016/j.dld.2005.03.011.

    Article  PubMed  Google Scholar 

  34. Lee DW, Li SAC, Ko CW, Chu DW, Chan KC, Poon CM, Sin KS, Leung KF, Sze TS, Chan AC, Chung SC: Use of a variable-stiffness colonoscope decreases the dose of patient-controlled sedation during colonoscopy: a randomized comparison of 3 colonoscopes. Gastrointest Endosc. 2007, 65: 424-429. 10.1016/j.gie.2006.08.027.

    Article  PubMed  Google Scholar 

  35. Sola-Vera J, Uceda F, Brotons A, Sáez J, Girona E, Pérez E, Picó MD, Grau C, Sillero C: Does the use of a variable stiffness colonoscope offer advantages during colonscopy under deep sedation?Results of a randomized trial. Eur J Gastroenterol Hepatol. 2011, 23 (7): 593-597. 10.1097/MEG.0b013e32834793d3.

    Article  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

No external funding was received for this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Huatian Gan.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Xie, Q., Chen, B., Liu, L. et al. Does the variable-stiffness colonoscope makes colonoscopy easier? A meta-analysis of the efficacy of the variable stiffness colonoscope compared with the standard adult colonoscope. BMC Gastroenterol 12, 151 (2012). https://doi.org/10.1186/1471-230X-12-151

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1471-230X-12-151

Keywords